Flame Photometers, Bomb Calorimeter DB-1142

  • Home
  • Flame Photometers, Bomb Calorimeter DB-1142
Flame Photometers, Bomb Calorimeter DB-1142
     
     
Flame Photometers, Bomb Calorimeter DB-1142

Flame Photometer Microprocessor Based

  • Price - : Rs.78,000
Technical Specification






MODE
General Serum Urine
Na: 0-100 ppm,
 Up to 250 meq/L, 1 : 100 dil
Na: 0-200 ppm
1:100 dil
Na:0-250 ppm
1:100 dil
K: 0-100 ppm
Up to 250 meq/L, 1 : 100 dil
K: 0-10 ppm
1:100 dil
K: 0-100 ppm
1:100 dil
Ca: 0-100 ppm
Up to 250 meq/L, 1 : 100 dil


 
Ca: 0-100 ppm
1:2 dil
Li: 0-100 ppm
Up to 250 meq/L, 1 : 100 dil
Li: 0-2 ppm
1:5 dil


 
Sensitivity General: Na : 0.5 ppm, K : 0.5 ppm, Li : 0.5 ppm, Ca: 15 ppm
Filters Na & K, Ca & Li (Optional if Ordered)
Resolution 0.1 ppm/ meq
Reproducibility < 1% coefficient of variation for 20 consecutive samples using 10 ppm, Na set as maximum standard.
Curve Fit Accuracy < 2% error when 3ppm Na/K and 5ppm Li, set as max.Standard.
Display 128 X 64 Graphical LCD
Average Time In built Software
Flame System LPG and Dry Oil free Air
Detector Highly Sensitive Silicon Photo Diode
Calibration Up to 5 point calibration with curve fitting Software.
Gas Control Adjustable with knob
Flame Failure Auto Detection
Gas Cut Off Gas feeding, auto cut off in case of Power Failure
Ignition System Auto Ignition
RS-232 Provision for Data Transfer to PC
Power 230 V +- 10%, 50 Hz. AC
Dimension 280X 250 X210 mm (L X W X H)
Weight 7.4 Kg (approx.)
Compressor Unit
Air Supply By oil free mini Compressor unit with pressure Regulator
Combustion LPG controlled by precision Regulator
Power 230 V +-10%, 50Hz AC
Dimension 290 X 250 X 210 mm (L X W X H)
Weight 8 Kg (approx.)

Download PDF